This is the metadata for: Warrawong_ANSTO_radon_2012-12-01_2013-02-28 (doi:10.1594/PANGAEA.871999) This file contains 3 months of hourly radon concentrations at Warrawong, NSW (-34.478, 150.884) provided by ANSTO in the context of the MUMBA campaign Species: radon Instrument: 1500L dual-flow-loop two-filter detector Owner: Australian Nuclear Science and Technology Organisation (ANSTO) Campaign: Instrument operational since 2006 - data provided in the context of the MUMBA campaign Location: Warrawong, NSW (-34.4781, 150.8843) Dates provided: 01-Dec-2012 until 28-Feb-2013 Original measurement time resolution: 30 minutes Averaging time: 1 hour Units: mBq/m3 Installation detail (inc. flow rate if relevant): The site is approximately 3 km from the NSW coast, at 40m above sea level, and sampling was conducted from 2m above ground level Data/sample analysis detail: radon concentrations are derived from the detector output by (1) scaling the raw counts by the detector sensitivity and subtracting instrumental background and (2) de-lagging the output by 30 minutes Calibration: monthly by injecting radon for 5 hours from a well-characterised PYLON source (19.58 +/- 4% kBq Ra-226) traceable to NIST standards Issues: none reported Corrections: These data have NOT been STP corrected Detection Limit: 40 mBq/m3 Bad or missing data reported as empty fields File headers: date yyyy-mm-ddTHH:MM:SSz radon (mBq/m3) mean hourly radon concentration in milliBecquerel per meter cubed Contact(s): Dr Scott Chambers szc@ansto.gov.au Dr Alastair Williams agw@ansto.gov.au